User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Electronic Research 2600S:Digital Timer DATE: 14-Dec-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 40 CONFIGURATION: HP 5335A STANDARD: Stanford Research FS700 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD * INITIAL CONDITIONS * 1.004 STD Stanford Research FS700 1.005 DISP Connect the UUT to the appropriate AC source and allow 1.005 DISP a 10 minute warm-up. 1.006 HEAD {* TIMEBASE ACCURACY TEST *} 1.007 DISP Connect the UUT 1MHz CLOCK output to the 5335A INPUT A. 1.008 IEEE [@5335] [TERM LF] 1.009 IEEE [@5335] IN[D2000]; 1.010 IEEE [@5335] FN1,AU1,AX1; 1.011 DISP Press the UUT START button. 1.012 IEEE [@5335] [D5000][I] 1.013 MEM/ 1000000 1.014 ACC 1.00000000MH 1P% 1.015 MEME 1.016 MEMC MH 0.000005U 2.001 HEAD {} 2.002 HEAD {* FUNCTION TEST *} 2.003 DISP Press the UUT {STOP button}. 2.004 EVAL {Did display stop counting and hold count?} 3.001 DISP Press the UUT {START button}. 3.002 EVAL {Did display start counting from held count?} 4.001 DISP Press the UUT STOP button. 4.002 DISP Press the UUT {RESET button}. 4.003 EVAL {Did display reset to all zeros?} 5.001 END